Electromagnetic flowmeter Promag 5P2B

Brand: Endress+Hauser

Electromagnetic flowmeter Promag 5P2B

Brand: Endress+Hauser

Forberg Smith, Endress+Hauser's exclusive USA representative in:
KY MI OH PA  Parts of WI WV 

Extended Extended
Heartbeat Technology Heartbeat Technology

Specs at a Glance

Wetted materials:
Liner: PFA; PTFE
Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum
Measuring range:
4 dm3/min to 1100 m3/h (1 to 4850 gal/min)
Max. process pressure:
PN 40, Class 300, 20K
Max. measurement error:
Volume flow: ±0.5 % o.r. ± 2 mm/s (0.08 in/s)
Medium temperature range:
Liner material PFA: –20 to +150 °C (–4 to +302 °F)
Liner material PTFE: –40 to +130 °C (–40 to +266 °F)

Field of application

Electromagnetic flowmeter
Inline version.
Dedicated to chemical and process
applications with corrosive liquids and
high medium temperatures.
Wide variety of electrode materials.
Liner: PTFE, PFA.
Loop-powered technology.
Robust dual-compartment housing.

Specifications

Industry

  • Chemical
  • Power & Energy
  • Oil & Gas

Product Family

  • Promag

Liquids

Product headline

  • The flowmeter for highest medium temperatures with genuine loop-powered technology.
  • Dedicated to chemical and process applications with corrosive liquids and high medium temperatures.

Sensor features

  • Diverse applications – wide variety of wetted materials. Energy‐saving flow measurement – no pressure loss due to cross section constriction. Maintenance‐free – no moving parts.
  • Nominal diameter: max. DN 200 (8"). All common Ex approvals. Liner made of PTFE or PFA.

Transmitter features

  • Convenient device wiring – separate connection compartment. Safe operation – no need to open the device due to display with touch control, background lighting. Integrated verification – Heartbeat Technology.
  • Loop-powered technology. Robust dual-compartment housing. Plant safety: worldwide approvals (SIL, Haz. area).

Nominal diameter range

  • DN 15 to 200 (½ to 8")

Wetted materials

  • Liner: PFA; PTFE
  • Electrodes: 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); Tantalum; Platinum

Measured variables

  • Volume flow, mass flow

Max. measurement error

  • Volume flow: ±0.5 % o.r. ± 2 mm/s (0.08 in/s)

Measuring range

  • 4 dm3/min to 1100 m3/h (1 to 4850 gal/min)

Max. process pressure

  • PN 40, Class 300, 20K

Medium temperature range

  • Liner material PFA: –20 to +150 °C (–4 to +302 °F)
  • Liner material PTFE: –40 to +130 °C (–40 to +266 °F)

Ambient temperature range

  • Flange material carbon steel: –10 to +60 °C (+14 to +140 °F)
  • Flange material stainless steel: –40 to +60 °C (–40 to +140 °F)

Sensor housing material

  • AlSi10Mg, coated

Transmitter housing material

  • AlSi10Mg, coated

Degree of protection

  • IP66/67, type 4X enclosure

Display/Operation

  • 4‐line backlit display with touch control (operation from outside)
  • Configuration via local display and operating tools possible
  • Remote display available

Outputs

  • 4‐20 mA HART (passive)
  • Pulse/frequency/switch output (passive)

Inputs

  • None

Digital communication

  • HART, PROFIBUS PA, FOUNDATION Fieldbus

Power supply

  • DC 18 to 35 V (4‐20 mA HART with/without pulse/frequency/switch output)

Hazardous area approvals

  • ATEX, IECEx, cCSAus, NEPSI, INMETRO, EAC, JPN, UK Ex

Other approvals and certificates

    Product safety

    • CE, C-Tick

    Functional safety

    • Functional safety according to IEC 61508, applicable in safety-relevant applications in accordance with IEC 61511

    Metrological approvals and certificates

    • Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
    • Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)

    Pressure approvals and certificates

    • CRN, PED

    Material certificates

    • 3.1 material

    Hygienic approvals and certificates

    • Drinking water approvals: ACS, NSF 61
    show more