Raman probe KRCAL

Brand: Endress+Hauser

Raman probe KRCAL

Brand: Endress+Hauser

Forberg Smith, Endress+Hauser's exclusive USA representative in:
KY MI OH PA  Parts of WI WV 

Extended Extended

Specs at a Glance

Spectral intensity calibration range:
CRS-785: 790.7 to 1074.5 nm
Spectral intensity reference:
Calibration Reference Standard

Field of application

Raman calibration and verification kits have everything needed to standardize your Raman bioprocess system for lab-to-process scalability. Kits are available for our multi optic (and single-use optic) systems, Raman flow assembly, and Rxn-46 probe, with others to come. Requiring no external power, these tools boost efficiency and reduce downtime via easy field deployment. These kits can be used without affecting sterilization in benchtop or single-use bioreactors or fermenters.

Cell culture: glucose, lactate, glutamine, amino acids, cell density, titer and more

Fermentation: glycerol, methanol, ethanol, sorbitol, biomass  and more

Downstream: aggregation, protein crystallization, formula stability, product CQA, protein concentration, buffer excipients

Midstream: analyte monitoring in permeate

Key Benefits

  • Uses modular hardware to simplify Raman analyzer standardization and verification, contained in a convenient, all-in-one kit
  • Facilitates easy calibration transfer of chemometric models for smooth lab-to-process scalability
  • Offers flexibility to calibrate anytime without impacting your process, which simplifies redundancy and mitigates risk
  • Enables quick instrument calibration and verification, reducing down time
  • Requires no external power for easier deployment in the field

Specifications

Industry

  • Life Science

Product Family

  • Raman

Liquid: Flow cell

Temperature

  • Operating:
  • 0 to 40 °C /32 to 104 °F
  • Storage:
  • -15 to 50 °C / 5 to 122 °F

Relative humidity

  • Storage:
  • < 80 % Non-Condensing

Unit dimensions (width x height x depth)

  • Kit: 235 x 192 x 85 mm
  • 9.3 x 7.6 x 3.4 inch

Weight

  • Kit: 1.5 kg / 3.3 lbs

Sampling probe compatibility

  • Raman Rxn-10 (with Raman flow bench)

Spectral intensity reference

  • Calibration Reference Standard

Spectral intensity output repeatability (at time of certification)

  • < ±2 %

Long term spectral uncertainty (at any wavelength)

  • ±6.05 %

Spectral intensity calibration range

  • CRS-785: 790.7 to 1074.5 nm

Liquid: Immersion

Temperature

  • Operating:
  • 0 to 40 °C /32 to 104 °F
  • Storage:
  • -15 to 50 °C / 5 to 122 °F

Relative humidity

  • Storage:
  • < 80 % Non-Condensing

Unit dimensions (width x height x depth)

  • Kit: 235 x 192 x 85 mm
  • 9.3 x 7.6 x 3.4 inch

Weight

  • Kit: 1.5 kg / 3.3 lbs

Sampling probe compatibility

  • Raman Rxn-10 (with bio multi optic)

Spectral intensity reference

  • Calibration Reference Standard

Spectral intensity output repeatability (at time of certification)

  • < ±2 %

Long term spectral uncertainty (at any wavelength)

  • ±6.05 %

Spectral intensity calibration range

  • CRS-785: 790.7 to 1074.5 nm

Liquid: Raman Rxn-46 probe

Temperature

  • Operating:
  • 0 to 40 °C /32 to 104 °F
  • Storage:
  • -15 to 50 °C / 5 to 122 °F

Relative humidity

  • Storage:
  • < 80 % Non-Condensing

Unit dimensions (width x height x depth)

  • Kit: 235 x 192 x 85 mm
  • 9.3 x 7.6 x 3.4 inch

Weight

  • Kit: 1.5 kg / 3.3 lbs

Sampling probe compatibility

  • Raman Rxn-46

Spectral intensity reference

  • Calibration Reference Standard

Spectral intensity output repeatability (at time of certification)

  • < ±2 %

Long term spectral uncertainty (at any wavelength)

  • ±6.05 %

Spectral intensity calibration range

  • CRS-785: 790.7 to 1074.5 nm
show more