Raman probe KRCAL
Brand: Endress+HauserRaman probe KRCAL
Brand: Endress+Hauser
Extended
Specs at a Glance
Field of application
Raman calibration and verification kits have everything needed to standardize your Raman bioprocess system for lab-to-process scalability. Kits are available for our multi optic (and single-use optic) systems, Raman flow assembly, and Rxn-46 probe, with others to come. Requiring no external power, these tools boost efficiency and reduce downtime via easy field deployment. These kits can be used without affecting sterilization in benchtop or single-use bioreactors or fermenters.
Cell culture: glucose, lactate, glutamine, amino acids, cell density, titer and more
Fermentation: glycerol, methanol, ethanol, sorbitol, biomass and more
Downstream: aggregation, protein crystallization, formula stability, product CQA, protein concentration, buffer excipients
Midstream: analyte monitoring in permeate
Key Benefits
- Uses modular hardware to simplify Raman analyzer standardization and verification, contained in a convenient, all-in-one kit
- Facilitates easy calibration transfer of chemometric models for smooth lab-to-process scalability
- Offers flexibility to calibrate anytime without impacting your process, which simplifies redundancy and mitigates risk
- Enables quick instrument calibration and verification, reducing down time
- Requires no external power for easier deployment in the field
Specifications
Brand
Industry
- Life Science
Product Family
- Raman
Liquid: Flow cell
Temperature
- Operating:
- 0 to 40 °C /32 to 104 °F
- Storage:
- -15 to 50 °C / 5 to 122 °F
Relative humidity
- Storage:
- < 80 % Non-Condensing
Unit dimensions (width x height x depth)
- Kit: 235 x 192 x 85 mm
- 9.3 x 7.6 x 3.4 inch
Weight
- Kit: 1.5 kg / 3.3 lbs
Sampling probe compatibility
- Raman Rxn-10 (with Raman flow bench)
Spectral intensity reference
- Calibration Reference Standard
Spectral intensity output repeatability (at time of certification)
- < ±2 %
Long term spectral uncertainty (at any wavelength)
- ±6.05 %
Spectral intensity calibration range
- CRS-785: 790.7 to 1074.5 nm
Liquid: Immersion
Temperature
- Operating:
- 0 to 40 °C /32 to 104 °F
- Storage:
- -15 to 50 °C / 5 to 122 °F
Relative humidity
- Storage:
- < 80 % Non-Condensing
Unit dimensions (width x height x depth)
- Kit: 235 x 192 x 85 mm
- 9.3 x 7.6 x 3.4 inch
Weight
- Kit: 1.5 kg / 3.3 lbs
Sampling probe compatibility
- Raman Rxn-10 (with bio multi optic)
Spectral intensity reference
- Calibration Reference Standard
Spectral intensity output repeatability (at time of certification)
- < ±2 %
Long term spectral uncertainty (at any wavelength)
- ±6.05 %
Spectral intensity calibration range
- CRS-785: 790.7 to 1074.5 nm
Liquid: Raman Rxn-46 probe
Temperature
- Operating:
- 0 to 40 °C /32 to 104 °F
- Storage:
- -15 to 50 °C / 5 to 122 °F
Relative humidity
- Storage:
- < 80 % Non-Condensing
Unit dimensions (width x height x depth)
- Kit: 235 x 192 x 85 mm
- 9.3 x 7.6 x 3.4 inch
Weight
- Kit: 1.5 kg / 3.3 lbs
Sampling probe compatibility
- Raman Rxn-46
Spectral intensity reference
- Calibration Reference Standard
Spectral intensity output repeatability (at time of certification)
- < ±2 %
Long term spectral uncertainty (at any wavelength)
- ±6.05 %
Spectral intensity calibration range
- CRS-785: 790.7 to 1074.5 nm