Prosonic Flow 9E1B
Brand: Endress+HauserProsonic Flow 9E1B
Brand: Endress+Hauser
Extended
Specs at a Glance
Measuring range:
0.025 to 5 m/s (0.02 to 16.4 ft/s)
0 to 6360 dm3/min (0 to 1680 gal/min)
0 to 6360 dm3/min (0 to 1680 gal/min)
Max. measurement error:
Volume flow:
- ± 0.5 % o.r. ± 0.02% o.f.s. for 0.5 to 10 m/s (1.6 to 33 ft/s)
- ± 0.07 % o.f.s. for flow velocities < 0.5 m/s (1.64 ft/s)
- ± 0.5 % o.r. ± 0.02% o.f.s. for 0.5 to 10 m/s (1.6 to 33 ft/s)
- ± 0.07 % o.f.s. for flow velocities < 0.5 m/s (1.64 ft/s)
Wetted materials:
Measuring tube: 1.4301 (F304)
Process Connection: 1.4571; 1.4404 (F316L); 1.0038 (S235JR); 1,4306 (F304L); 1.4307 (F304L); A105
Process Connection: 1.4571; 1.4404 (F316L); 1.0038 (S235JR); 1,4306 (F304L); 1.4307 (F304L); A105
Max. process pressure:
PN 25 / ASME Cl. 150
Medium temperature range:
0 to 150° (17.8 to 302 °F)
Field of application
Measuring of demineralized water in utilities, for example in boiler condensate return lines, requires reliable devices. Prosonic Flow E 100 provides economical and accurate bidirectional flow measurement across all industries. With its integrated temperature measurement, Prosonic Flow E 100 proves to be a multivariable flowmeter for considerable additional value in your plant. Heartbeat Technology enables measurement reliability and extension of recalibration cycles.
The measuring principle is independent of pressure, density, temperature and conductivity
Bidirectional measuring of demineralized water applications for utilities, e.g. boiler condensate return lines
Device properties:
- Accuracy up to ±0.5 % (flow) or acc. to EN 1434 Cl. 2, ±2.0 °C (± 3.6 °F) (temp.)
- Process temperatures up to 150 °C (302 °F)
- Entire meter housing made of stainless steel
- 4-20 mA HART, pulse/frequency output
- Local display for reading and monitoring available
- Robust transmitter housing
Key Benefits
- Long-term stability – reliable, robust sensor
- Reducing further measuring point – multivariable device
- Dependable flow measurement – high turndown (200:1)
- Time-saving local operation without additional software and hardware – integrated web server
- Extended calibration intervals – integrated device verification due to Heartbeat Technology
- Easy commissioning – brief parameter explanations
Specifications
Brand
Product Family
- Prosonic
Liquids
Product headline
- Economical Ultrasonic flowmeter with integrated temperature measurement.
- Bidirectional measuring of demineralized water applications for Utilities, e.g. boiler condensate return lines.
Sensor features
- Long-term stability – reliable, robust sensor. Reducing further measuring point – multivariable device. Dependable flow measurement – high turndown (200:1).
- Measurement accuracy up to ±0.5 % (flow) or according to EN 1434 Cl. 2, ±2.0 °C (± 3.6 °F) (temperature). Process temperatures up to 150 °C (302 °F). Entire meter housing made of stainless steel.
Transmitter features
- Time-saving local operation without additional software and hardware – integrated web server. Extended calibration intervals – integrated device verification due to Heartbeat Technology. Easy commissioning – brief parameter explanations.
- 4-20 mA HART, pulse/frequency output. Local display for reading and monitoring available. Robust transmitter housing.
Nominal diameter range
- DN 50 to 150 (2"to 6")
Wetted materials
- Measuring tube: 1.4301 (F304)
- Process Connection: 1.4571; 1.4404 (F316L); 1.0038 (S235JR); 1,4306 (F304L); 1.4307 (F304L); A105
Measured variables
- Volume flow, calculated mass flow, flow velocity, sound velocity, temperature
Max. measurement error
- Volume flow:
- - ± 0.5 % o.r. ± 0.02% o.f.s. for 0.5 to 10 m/s (1.6 to 33 ft/s)
- - ± 0.07 % o.f.s. for flow velocities < 0.5 m/s (1.64 ft/s)
Measuring range
- 0.025 to 5 m/s (0.02 to 16.4 ft/s)
- 0 to 6360 dm3/min (0 to 1680 gal/min)
Max. process pressure
- PN 25 / ASME Cl. 150
Medium temperature range
- 0 to 150° (17.8 to 302 °F)
Ambient temperature range
- -20 to 60 °C (-4 to 140 °F)
Sensor housing material
- Stainless steel
Transmitter housing material
- Compact:
- AlSi10Mg, coated
- Stainless steel, 1.4301
Degree of protection
- Standard: IP66/67, Type 4X enclosure
- With opened housing: IP20, Type 1 enclosure
Display/Operation
- 4‐line backlit display available (no local operation)
- Configuration via web browser and operating tools possible
Outputs
- 4‐20 mA HART (active)
- Pulse/frequency/switch output (passive)
Inputs
- None
Power supply
- DC 19.2 to 28.8 V
Hazardous area approvals
- None
Product safety
- CE, C-Tick, EAC marking
Metrological approvals and certificates
- Calibration performed on accredited calibration facilities (acc. to ISO/IEC 17025)
- Heartbeat Technology complies with the requirements for measurement traceability according to ISO 9001:2015 – Section 7.1.5.2 a (TÜV SÜD attestation)
- Calibration according to EN 1434 Class 2
Pressure approvals and certificates
- PED
Material certificates
- 3.1 material