Raman Rxn KRXN2B

Brand: Endress+Hauser

Raman Rxn KRXN2B

Brand: Endress+Hauser

Forberg Smith, Endress+Hauser's exclusive USA representative in:
KY MI OH PA  Parts of WI WV 

Extended Extended

Specs at a Glance

Spectral coverage:
Starter 785 nm: 300-3300 cm-1
Base model 532 nm: 100-4350* cm-1
Base model 785 nm: 100-3425* cm-1
Base model 1000 nm: 200-2400 cm-1
Hybrid 785 nm: 150-1890** cm-1

*Low signal-to-noise ratio at <150 cm-1
**Low signal-to-noise ratio at <175 cm-1
Laser wavelength:
Starter: 785 nm

Base Model: 532 nm, 785 nm, 1000 nm

Hybrid: 785 nm

Field of application

Adeptly harness the power of Raman spectroscopy with the Raman Rxn2 analyzer. Designed for use in analytical laboratories with model transfer capabilities, the Raman Rxn2 is heavily relied on for routine sample identification, support of R&D projects, early process development, and scale-up settings for in situ analysis. Available as a benchtop or on a mobile wheeled cart, the Raman Rxn2 offers location convenience and portability for process development laboratories.

Chemical: polymerization, extrusion monitoring, polymer blending, catalysis investigation

Pharmaceutical: reaction chemistry, yield crystallization, polymorph, blending, drying, granulations

Biopharmaceutical: bioprocess monitoring and control, PAT/QbD applications, cell culture, fermentation, downstream operations, cGLP/cGMP compatible

Food & Beverage: cultured meat, fermentation, oil or fat quality

Key Benefits

  • Reliable real-time, in situ measurements
  • Intuitive, embedded control software via touchscreen or remote interface
  • Convenience of a single base unit supporting up to four probes
  • Sequential operation for fast analysis per channel and programmable channel interrogation
  • Converts acquired Raman spectra into process knowledge using built-in multivariate predictors
  • Suitable for outputs into hazardous area/classified environments
  • Proven model transferability to process development and manufacturing scales

Specifications

Industry

  • Chemical
  • Food & Beverage
  • Oil & Gas
  • Life Science

Product Family

  • Raman

Gases

Laser wavelength

  • 532 nm

Spectral coverage

  • 532 nm: 100-4350* cm-1
  • *Low signal-to-noise ratio at <150 cm-1

Spectral resolution

  • 5 cm-1 (532 nm)

Channels

  • Up to four channels

Ambient temperature

  • Operating:
  • 15 to 30 °C
  • 59 to 86 °F
  • Storage:
  • -15 to 50 °C
  • 5 to 122 °F

Relative humidity

  • 20 to 80 %, non-condensing

Input voltage

  • 100-240 V, 50-60 Hz, ±10%

Power consumption (W)

  • 400 (max)
  • 250 (typical start-up)
  • 120 (typical running)

Warm up time (minutes)

  • 120

Unit dimensions (width x height x depth)

  • Benchtop model:
  • 279 x 483 x 592 mm
  • 10.99 x 19.02 x 23.31 inch
  • Cart model:
  • 685 x 1022 (to tabletop) x 753 mm
  • 26.97 x 40.24 (to tabletop) x 29.65 inch

Weight

  • Base model:
  • 32 kg / 71 lb
  • Cart model:
  • 93 kg / 205 lb

Sampling probe compatibility

  • Raman Rxn-10 with accessory optics
  • Raman Rxn-40
  • Raman Rxn-41

Automation interface

  • OPC
  • Modbus
  • HTTPS
  • (contact us for other options)

Installation options

  • Benchtop (standard) or mobile wheeled cart

Hazardous area certifications

  • ATEX, CSA, IECEx, UKCA, JPEx

Liquid

Laser wavelength

  • Starter: 785 nm
  • Base Model: 532 nm, 785 nm, 1000 nm
  • Hybrid: 785 nm

Spectral coverage

  • Starter 785 nm: 300-3300 cm-1
  • Base model 532 nm: 100-4350* cm-1
  • Base model 785 nm: 100-3425* cm-1
  • Base model 1000 nm: 200-2400 cm-1
  • Hybrid 785 nm: 150-1890** cm-1
  • *Low signal-to-noise ratio at <150 cm-1
  • **Low signal-to-noise ratio at <175 cm-1

Spectral resolution

  • Starter:
  • 10 cm-1 (average)
  • Base model (average):
  • 5 cm-1 (532 nm)
  • 4 cm-1 (785 nm)
  • 6 cm-1 (1000 nm)
  • Hybrid:
  • 4 cm-1 (785 nm) average

Channels

  • Starter:
  • Single channel
  • Base Model:
  • Up to four channels
  • Hybrid:
  • Up to two channels

Ambient temperature

  • Operating:
  • 15 to 30 °C
  • 59 to 86 °F
  • Storage:
  • -15 to 50 °C
  • 5 to 122 °F

Relative humidity

  • 20 to 80 %, non-condensing

Input voltage

  • 100-240 V, 50-60 Hz, ±10%

Power consumption (W)

  • 400 (max)
  • 250 (typical start-up)
  • 120 (typical running)

Warm up time (minutes)

  • 120

Unit dimensions (width x height x depth)

  • Benchtop model:
  • 279 x 483 x 592 mm
  • 10.99 x 19.02 x 23.31 inch
  • Cart model:
  • 685 x 1022 (to tabletop) x 753 mm
  • 26.97 x 40.24 (to tabletop) x 29.65 inch

Weight

  • Base model:
  • 32 kg / 71 lb
  • Cart model:
  • 93 kg / 205 lb

Sampling probe compatibility

  • Starter:
  • Raman Rxn-10 (with accessory optics)
  • Base model:
  • Raman Rxn-10 (with accessory optics), Rxn-40, Rxn-41, Rxn-45, Rxn-46
  • Hybrid:
  • Channel 1 - Raman Rxn-20 (with accessory optics)
  • Channel 2 - Raman Rxn-10 (with accessory optics), Rxn-40, Rxn-41, Rxn-45, Rxn-46

Automation interface

  • OPC
  • Modbus
  • HTTPS
  • (contact us for other options)

Installation options

  • Benchtop (standard) or mobile wheeled cart

Hazardous area certifications

  • ATEX, CSA, IECEx, UKCA, JPEx

Solids

Laser wavelength

  • Starter: 785 nm
  • Base Model: 532 nm, 785 nm, 1000 nm
  • Hybrid: 785 nm

Spectral coverage

  • Starter 785 nm: 300-3300 cm-1
  • Base model 532 nm: 100-4350* cm-1
  • Base model 785 nm: 100-3425* cm-1
  • Base model 1000 nm: 200-2400 cm-1
  • Hybrid 785 nm: 150-1890** cm-1
  • *Low signal-to-noise ratio at <150 cm-1
  • **Low signal-to-noise ratio at <175 cm-1

Spectral resolution

  • Starter:
  • 10 cm-1 (average)
  • Base model (average):
  • 5 cm-1 (532 nm)
  • 4 cm-1 (785 nm)
  • 6 cm-1 (1000 nm)
  • Hybrid:
  • 4 cm-1 (785 nm) average

Channels

  • Starter:
  • Single channel
  • Base Model:
  • Up to four channels
  • Hybrid:
  • Up to two channels

Ambient temperature

  • Operating:
  • 15 to 30 °C
  • 59 to 86 °F
  • Storage:
  • -15 to 50 °C
  • 5 to 122 °F

Relative humidity

  • 20 to 80 %, non-condensing

Input voltage

  • 100-240 V, 50-60 Hz, ±10%

Power consumption (W)

  • 400 (max)
  • 250 (typical start-up)
  • 120 (typical running)

Warm up time (minutes)

  • 120

Unit dimensions (width x height x depth)

  • Benchtop model:
  • 279 x 483 x 592 mm
  • 10.99 x 19.02 x 23.31 inch
  • Cart model:
  • 685 x 1022 (to tabletop) x 753 mm
  • 26.97 x 40.24 (to tabletop) x 29.65 inch

Weight

  • Base model:
  • 32 kg / 71 lb
  • Cart model:
  • 93 kg / 205 lb

Sampling probe compatibility

  • Starter:
  • Raman Rxn-10 (with accessory optics)
  • Base model:
  • Hybrid:
  • Channel 1 - Raman Rxn-20 (with accessory optics)
  • Channel 2 - Raman Rxn-10 (with accessory optics)

Automation interface

  • OPC
  • Modbus
  • HTTPS
  • (contact us for other options)

Installation options

  • Benchtop (standard) or mobile wheeled cart

Hazardous area certifications

  • ATEX, CSA, IECEx, UKCA, JPEx
show more