EngyCal RH33
Brand: Endress+HauserEngyCal RH33
Brand: Endress+Hauser
Lean Specs at a Glance
Output:
1x 4...20mA
2x digital (Open Collector)
2x digital (Open Collector)
Input:
1x Puls/Analog
2x RTD/Analog
Loop power supply 24V DC (+/-16%)
2x RTD/Analog
Loop power supply 24V DC (+/-16%)
Display:
160 x 80 Dot-Matrix LCD with white backlit
colour change in case of alarm event
active display area 70 x 34 mm
colour change in case of alarm event
active display area 70 x 34 mm
Calculations:
heat quantity and heat quantity difference
Field of application
The BTU meter RH33 is used in applications with liquid energy carriers.It calculates the thermal energy of water acc. to EN1434, glycol/water mixtures or other fluids such as thermal oils. Temperature sensor matching using calibrated temperature sensors is done in the device.Potential for cost savings can be shown by using the software available as an accessory.The device has a custody transfer approval and allows bi-directional measurement, e.g. charging/discharging of a heat accumulator.
Recording and billing the quantities of heat and cold in:
- Heating circuits
- Cooling circuits
- Combined heating and cooling circuits
- Typical application areas in industry, long-distance heat networks and building automation.
Key Benefits
- Transparent energy consumption helps you save on energy costs
- Calibrated, electronically paired temperature sensors ensure the highest accuracy and enable replacement of individual temperature sensors even for certified devices in the field (without reapproval!)
- Tariff counter for requirements-based billing
- Detailed data logging of current and counter values and of error messages, off-limit conditions and changes to operating parameters
- Standard models are suitable for connecting and supplying all common volume flow transmitters and temperature measuring points
- Remote readout via Ethernet and fieldbuses
- Verified and certified reliability and accuracy
Specifications
Brand
Industry
- Chemical
- Water, Wastewater & Waste
- Food & Beverage
- Power & Energy
- Mining, Minerals & Metals
- Oil & Gas
- Pulp & Paper (incl. Printing)
- Other industries
- Life Science
Product Family
- EngyCal
Energy & Application Manager
Measuring principle
- Energy manager
Function
- Monitoring and billing of energy in heating and cooling applications as well as combined heating-and cooling circles.
- Typical applications can be found in industries, district heating and building automation.
Calculations
- heat quantity and heat quantity difference
Number of applications
- 1
Calculation standards
- IAPWS-97
Communication
- web server
- USB
- Ethernet
- Modbus RTU/TCP Slave
- M-Bus
Power supply
- Not defined
Loop power supply
- Low voltage power supply:
- 100 bis 230 V AC (-15% / +10%)
- Small voltage power supply:
- 24 V DC (-50% / +75%)
- 24 V AC (±50%)
Protection class
- IP65
Input
- 1x Puls/Analog
- 2x RTD/Analog
- Loop power supply 24V DC (+/-16%)
Output
- 1x 4...20mA
- 2x digital (Open Collector)
Dimensions (WxHxD)
- 144 x 144 x 103.1 mm (5.67" x 5.67" x 4.06")
Operation
- 3 button on site or via FieldCare
- read out of historical / logged data via Field Data Manager Software (SQL database and visualization interface)
- selectable language
Display
- 160 x 80 Dot-Matrix LCD with white backlit
- colour change in case of alarm event
- active display area 70 x 34 mm
Software functions
- Calculated Values:
- Energy, volume, density, enthalpy & enthalpy difference, DP-Flow-Compensation, mass, temperature difference
- Counters:
- Volume, mass, energy, counter in case of failure
- Optional:
- tariff 1, tariff 2 or seperated heating energy, cooling energy, bilance energy
Certificates
- CSA GP
- MID 004 (custody transfer) according EN1434 (water/other liquids)
- OIML R75
Thermal Energy Measurement
Function
- Monitoring and billing of energy in heating and cooling applications as well as combined heating-and cooling circles.
- Typical applications can be found in industries, district heating and building automation.
Certification
- CE, UL listed, CSA GP, MID 004 (custody transfer) according EN1434 (water/other liquids) and OIML R75
Input
- 1x Puls/Analog,
- 2x RTD/Analog,
- Loop power supply 24V DC (+/-16%)
Output
- 1x 4...20mA,
- 2x digital (Open Collector)
Relay output
- 2x
Auxiliary power supply
- Low voltage power supply: 100 bis 230 V AC (-15% / +10%); Small voltage power supply: 24 V DC (-50% / +75%), 24 V AC (±50%)
Dimensions (wxhxd)
- 144 x 144 x 103.1 mm
- (5.67" x 5.67" x 4.06")
Software
- Calculated Values:
- Energy, volume, density, enthalpy & enthalpy difference,
- DP-Flow-Compensation,
- mass,
- temperature difference;
- Counters: volume, mass, energy, counter in case of failure
- Optional: tariff 1, tariff 2 or seperated heating energy, cooling energy, bilance energy
Operation
- 3 button on site or via FieldCare; read out of historical / logged data via Field Data Manager Software (SQL database and visualization interface), selectable language
Display
- 160 x 80 Dot-Matrix LCD with white backlit, colour change in case of alarm event, active display area 70 x 34 mm
Others
- electronic temperature sensor matching via CvD coefficients, logbook measured values, custody transfer logbook, event logbook, limit monitoring
Calculations
- IAPWS-IF97