Raman probe KR20
Brand: Endress+HauserRaman probe KR20
Brand: Endress+Hauser
Extended
Specs at a Glance
Window: optical-grade materials
Field of application
For solids and semi-solids analysis, the Raman Rxn-20 probe reigns supreme. Its state-of-the-art, stainless-steel probe design produces representative, focus-free, quantitative Raman measurements. The Rxn-20 probe illuminates a large surface area and eliminates the need to align the probe for surface roughness. It provides heterogenous solids sampling via volumetric analysis, which is highly useful for measuring tablets, capsules, food solids, and polymer beads.
Chemicals: final product quality, blend impurities, crystallinity, raw materials
Polymers: extruded pellet quality, crystallinity, density, raw materials
Pharmaceutical: crystallinity, polymorphism, granulation, blend uniformity, content uniformity, coating, tableting
Food & Beverage: quality of dairy solids, meat and fish composition
Key Benefits
- Non-contact measurement of heterogeneous solids for better representation
- Improved process control and efficiency through faster measurements
- Non-destructive measurements from a distance
- Reproducible sampling
- Sampling flexibility with a variety of focus-free and immersion Rxn-20 accessory optics
- No need to align probe for surface roughness
- Surface and deep layer (volumetric) analysis
Specifications
Brand
Industry
- Chemical
- Food & Beverage
- Life Science
Product Family
- Raman
Solids
Laser wavelength
- 785 nm
Spectral coverage
- 175-1890 cm-1 (785 nm)
Ambient temperature
- 10 to 40 °C / 50 to 104 °F
Relative humidity
- 20 to 80 %, non-condensing
Maximum laser power into probehead (mW)
- < 499
Sample interface
- Temperature:
- 10 to 40 °C / 50 to 104 °F
- Relative humidity:
- 20 to 80 %, non-condensing
Pressure
- Ambient
Body and window materials
- Body: 316L stainless steel
- Window: optical-grade materials
Fiber optic cable
- Connections:
- proprietary electro-optic
- Temperature:
- -40 to 70 °C / -40 to 158 °F
- Lengths available:
- 3000 mm / 118.12 inch
- 10000 mm / 393.71 inch
- 15000 mm /590.56 inch
- Minimum bend radius:
- 75 mm / 2.96 inch
Length
- Rxn-20 probe (including fiber cable bend radius):
- 312 mm / 12.29 inch
Diameter
- Raman Rxn-20:
- 48 mm / 1.89 inch
Working distance
- Based on the sampling optic selected
Hazardous area certifications
- ATEX, CSA, IECEx, UKCA, JPEx