Raman Rxn KRXN5C

Brand: Endress+Hauser

Raman Rxn KRXN5C

Brand: Endress+Hauser

Forberg Smith, Endress+Hauser's exclusive USA representative in:
KY MI OH PA  Parts of WI WV 

Extended Extended

Specs at a Glance

Spectral coverage:
150-3425 cm-1 (532 nm)
Laser wavelength:
532 nm

Field of application

The Raman Rxn5 process analyzer is an industry-leading, turn-key, laser-based analyzer powered by Kaiser Raman technology. It provides quantitative chemical composition measurements using Raman spectroscopy. The design of the Raman Rxn5 incorporates customer requirements for serviceability and hazard-area certification, provides a compact footprint, and minimizes utilities consumption. The high optical efficiency of the analyzer makes it ideal for typical gas-phase sample measurements.

Process measurement points:

  • Steam methane reformers
  • Gasifiers (coal, petcoke, waste, biomass)
  • Shift converters
  • Hydrotreaters
  • Hydrocrackers
  • CO2 removal
  • Finished product synthesis loops
  • Typical industries:
  • Hydrogen production/purity, HyCO, Hydrogen recycle,
  • Methanol production
  • Ammonia production
  • Gas turbine fuel feed
  • Synthetic natural gal/IGCC power plants

Key Benefits

  • Non-destructive analysis of gases, including homonuclear diatomics (H2, N2, O2)
  • Compact, smaller than most wall-mount gas chromatographs (GCs)
  • Can replace up to four traditional analyzers with a single base unit and four probes
  • Rugged with minimal utility requirements, no sample transport, and no consumables (columns, valves, gases, pumps)
  • Minimal analyzer technician time with maximum technician safety (no toxic gases are transported to the analyzer)
  • Measures four streams simultaneously for faster multi-channel inputs
  • -20 to +50 °C operating environment

Specifications

Industry

  • Power & Energy
  • Oil & Gas

Product Family

  • Raman

Gases

Laser wavelength

  • 532 nm

Spectral coverage

  • 150-3425 cm-1 (532 nm)

Spectral resolution

  • 5 cm-1 (532 nm)

Channels

  • Up to four channels

Ambient temperature

  • Operating:
  • -20 to 50 °C / -4 to 122 °F
  • Recommended storage:
  • -30 to 60 °C / -22 to 140 °F

Relative humidity

  • 0 to 90 %, non-condensing

Input voltage

  • 90-264 VAC, 47-63 Hz standard

Power consumption (W)

  • <300 (max)
  • <300 (typical start-up)
  • <200 (typical running)

Warm up time (minutes)

  • 120

Unit dimensions (width x height x depth)

  • 457 x 834 x 254 mm
  • 18.00 x 32.84 x 10.00 inch

Weight

  • 61.2 kg / 135 lb

Sampling probe compatibility

  • Raman Rxn-30

Automation interface

  • Modbus (TCP/IP or RS485)

Installation options

  • IP 56 wall-mounted, sealed/purged enclosure

Hazardous area certifications

  • ATEX, CSA, IECEx, UKCA, JPEx, KCs
show more