Raman Rxn KRXN5C
Brand: Endress+HauserRaman Rxn KRXN5C
Brand: Endress+Hauser
Extended
Specs at a Glance
Spectral coverage:
150-3425 cm-1 (532 nm)
Laser wavelength:
532 nm
Field of application
The Raman Rxn5 process analyzer is an industry-leading, turn-key, laser-based analyzer powered by Kaiser Raman technology. It provides quantitative chemical composition measurements using Raman spectroscopy. The design of the Raman Rxn5 incorporates customer requirements for serviceability and hazard-area certification, provides a compact footprint, and minimizes utilities consumption. The high optical efficiency of the analyzer makes it ideal for typical gas-phase sample measurements.
Process measurement points:
- Steam methane reformers
- Gasifiers (coal, petcoke, waste, biomass)
- Shift converters
- Hydrotreaters
- Hydrocrackers
- CO2 removal
- Finished product synthesis loops
- Typical industries:
- Hydrogen production/purity, HyCO, Hydrogen recycle,
- Methanol production
- Ammonia production
- Gas turbine fuel feed
- Synthetic natural gal/IGCC power plants
Key Benefits
- Non-destructive analysis of gases, including homonuclear diatomics (H2, N2, O2)
- Compact, smaller than most wall-mount gas chromatographs (GCs)
- Can replace up to four traditional analyzers with a single base unit and four probes
- Rugged with minimal utility requirements, no sample transport, and no consumables (columns, valves, gases, pumps)
- Minimal analyzer technician time with maximum technician safety (no toxic gases are transported to the analyzer)
- Measures four streams simultaneously for faster multi-channel inputs
- -20 to +50 °C operating environment
Specifications
Brand
Industry
- Power & Energy
- Oil & Gas
Product Family
- Raman
Gases
Laser wavelength
- 532 nm
Spectral coverage
- 150-3425 cm-1 (532 nm)
Spectral resolution
- 5 cm-1 (532 nm)
Channels
- Up to four channels
Ambient temperature
- Operating:
- -20 to 50 °C / -4 to 122 °F
- Recommended storage:
- -30 to 60 °C / -22 to 140 °F
Relative humidity
- 0 to 90 %, non-condensing
Input voltage
- 90-264 VAC, 47-63 Hz standard
Power consumption (W)
- <300 (max)
- <300 (typical start-up)
- <200 (typical running)
Warm up time (minutes)
- 120
Unit dimensions (width x height x depth)
- 457 x 834 x 254 mm
- 18.00 x 32.84 x 10.00 inch
Weight
- 61.2 kg / 135 lb
Sampling probe compatibility
- Raman Rxn-30
Automation interface
- Modbus (TCP/IP or RS485)
Installation options
- IP 56 wall-mounted, sealed/purged enclosure
Hazardous area certifications
- ATEX, CSA, IECEx, UKCA, JPEx, KCs