Raman probe KR45
Brand: Endress+HauserRaman probe KR45
Brand: Endress+Hauser
Extended
Specs at a Glance
Wetted materials:
Body: 316L stainless steel
Window: Proprietary material, optimized for bioprocesses
Process connection: PG13.5 for industry standard sensor housings, welded port connectors available
Surface finish: Ra 15 with electropolish
Adhesive: USP Class VI and ISO993 compatible
Window: Proprietary material, optimized for bioprocesses
Process connection: PG13.5 for industry standard sensor housings, welded port connectors available
Surface finish: Ra 15 with electropolish
Adhesive: USP Class VI and ISO993 compatible
Sterilization method:
CIP/SIP
Laser wavelength:
785 nm, 1000 nm
Field of application
The Raman Rxn-45 probe capitalizes on the power of Raman in bioprocess manufacturing by measuring multiple, specific components in real-time for continuous, around-the-clock process feedback. It also meets the daunting sampling requirements of compliance, sterilization, port compatibility, and convenience. The Raman Rxn-45 probe was designed for installation in development and cGMP stainless-steel reactors and is successfully used to provide an analytical “eye” into large-scale bioprocesses.
Cell culture: glucose, lactate, amino acids, cell density, titer, and more
Fermentation: glucose, glycerol, acetate, methanol, ethanol, biomass, and more
Key Benefits
- Measures multiple components in real-time for automated 24/7 process feedback
- Provides long-term measurement stability
- Offers a suitable surface finish for cGMP manufacturing
- Provides compatibility with industry standard bioreactor side ports and sensor housings
- Offers the flexibility of being installed in development and production reactors
- Reduces sterilization and cleaning burdens with CIP/SIP standard compatibility
Specifications
Brand
Industry
- Life Science
Product Family
- Raman
Liquids
Laser wavelength
- 785 nm, 1000 nm
Spectral coverage
- Probe spectral coverage is limited by the coverage of the analyzer being used.
Maximum laser power into probehead (mW)
- < 499
Sample interface
- Temperature:
- -30 to 150 °C / -22 to 302 °F
- Temperature ramp:
- ≤ 30 °C/min
- ≤ 54 °F/min
Pressure
- Max pressure:
- 13.8 barg / 200 psig
Wetted materials
- Body: 316L stainless steel
- Window: Proprietary material, optimized for bioprocesses
- Process connection: PG13.5 for industry standard sensor housings, welded port connectors available
- Surface finish: Ra 15 with electropolish
- Adhesive: USP Class VI and ISO993 compatible
Fiber optic cable
- Cable sold separately
Length
- 120 mm / 4.73 inch
Diameter
- 12 mm / 0.48 inch
Sterilization method
- CIP/SIP