Liquicap FMI51/52
Brand: Endress+HauserLiquicap FMI51/52
Brand: Endress+Hauser
Extended
Specs at a Glance
Max. measurement distance:
0.1 m ... 4.0 m
(0.3 ft ... 13 ft)
(0.3 ft ... 13 ft)
Accuracy:
Repeatablity 0,1%
Process pressure / max. overpressure limit:
Vacuum ... 100 bar
(Vaccum ... 1450 psi)
(Vaccum ... 1450 psi)
Process temperature:
-80°C...200°C
-112°F...392°F
-112°F...392°F
Main wetted parts:
Insulation material: PTFE,PFA
316L
316L
Field of application
The Liquicap FMI51 is a reliable fully insulated rod probe for continuous level monitoring in liquids, particularly in build-up forming media and extremely high temperatures. The measurement is independent of the dielectric constant (dc).
Specifications
Brand
Industry
- Chemical
- Water, Wastewater & Waste
- Food & Beverage
- Power & Energy
- Mining, Minerals & Metals
- Oil & Gas
- Pulp & Paper (incl. Printing)
- Life Science
Product Family
- Liquicap
Continuous / Liquids
Characteristic / Application
- Fully insulated rod probe, for standart- and extreme process conditions (temperature, pressure, build-up)
Interface measurement
- Interfaces liquid / liquid also with emulsion layers
Specialities
- Inactive length
- Factory calibrated
- short response time by change of value
Supply / Communication
- 12-36V DC HART
- PFM
Accuracy
- Repeatablity 0,1%
Linearity error for conductive liquids
- <0,25%
Ambient temperature
- -50°C...+70°C
- -58°F...+158°F
Process temperature
- -80°C...200°C
- -112°F...392°F
Process pressure / max. overpressure limit
- Vacuum ... 100 bar
- (Vaccum ... 1450 psi)
Main wetted parts
- Insulation material: PTFE,PFA
- 316L
Process connection
- G1/2, G 3/4, G 1, G1 1/2 /NPT 1/2, NPT 3/4, NPT 1", NPT1 1/2
- Flanges from DN25.../ASME 1".../JIS...
Process connection hygienic
- Tri-Clamp ISO02852 gap free plated
- Dairy coupling
- Flush-mounted seal
Sensor length
- Total length: 6m (20ft)
- Inactive length: max. 2m (7ft)
Max. measurement distance
- 0.1 m ... 4.0 m
- (0.3 ft ... 13 ft)
Communication
- 4...20mA HART
- PFM
Certificates / Approvals
- ATEX, FM, CSA, IEC Ex, TIIS, INMETRO, NEPSI, EAC
Safety approvals
- SIL
Design approvals
- EN 10204-3.1
- NACE MR0175
Hygienic approvals
- 3A, EHEDG
Marine approval
- GL/ ABS/ DNV
Options
- Separate housing
- gas-tight probe seal
Application limits
- Insufficient clearance towards ceiling
- Changing, non-conductive
- media, conductivity
- < 100 μS/cm
- Notice the pressure and temperature derating